CLINICAL CHEMISTRY ANALYSER
(AUTOLYSER)
MODELS:
without 4-steps washing
station and without cooled reagent tray
with 4-steps washing station
and cooled reagent tray
TECHNICAL
SPECIFICATIONS
Sampling and reagent center:
48 positions sample carrousel Primary tubes
with barcode option Accepts pediatric micro-samples Sample volume:
programmable 2-100µl
48 cooled positions reagent caroussel Reagent
volume: programmable: Reagent 1: from 0 to 1600µl Reagent 2: from 0
to 1600µl Typical reagent volume: 350µl
Pipetting
system:
- Reagent probe heater
- Capacitive liquid level sensor
- Washing system outside and inside the probe
Reaction
tray:
- 80 positions reaction tray with washing station
- 4 steps washing station
- Water consumption: 1 liter / 500 tests
- Warmed air incubator: ambient, 30°C and
37°C
Optic system
- Double beam
- Interference filters: 340, 405, 450, 505, 550, 590, 650, 700 and 750nm
- Bandwidth: 10nm
- Photometric range: -0.1 to 3.6A
- Light source: halogen lamp 6V-20W
Analysis
modes:
- End point with sample blank or reagent blank
- Use of factor or standard
- Priority selection by sample (profile) or by reagent (blank)
- Use of calibration curve with up to 10 standards
- Automatic curve fit
- Coagulation times by turbidimetry
- Fast and two-point kinetics (zero and first order)
- Routine, batch, STAT procedures, profiles
- Automatic analysis redilution on abnormal levels,
- excessive substrate
- consumption and / or lack of linearity
- Full quality control: Levy-Jennings plots, Westgard rules
- Import / export of data, methods and historic file
- Automatic backup procedure
Throughput:
- Up to 230 tests / hour depending on configuration
Power
requirements:
- 110/230 VAC; 50/60 Hz; 100VA
Data
management:
- Computer: PentiumTM or last state of the art computer
- Color SVAG monitor
- Multitasking Windows software, Mouse
- 3.5 1/2” disk drive and CD-Rom drive
- Printer: 80 columns.
- Interface: bi-directional RC 232C
Dimensions /
Weight
|