Features: - Discrete, random access, fully automated
- 400 tests per hour, up to 640 tests per hour with ISE
- Up to77 on-board chemistries & 4 ions
- Refrigerated reagent compartment
- Onboard capacity of 90 sample positions
- Automatic probe cleaning, liquid level detection & collision
protection
- 8-step auto washing system
- 12 wavelength: 340-800nm
- Automatic sample dilution for abnormal sample
- Internal bar code reader (optional)
- Bidirectional connection to LIS host
Technical Specifications:
System function:
- Automatic, Discrete, Random Access STAT sample priority
- Throughput: 400 tests/hour, up to 640 tests/hour with ISE
- Measuring principles: Absorbance photometry, Turbidimetry
-Methodology: End-point, Initial rate, Kinetic, optional ISE, Up to
4-reagent chemistries, monochromatic / bichromatic
- Programming: User defined profiles and calculation chemistries
Sample Handling: - Sample tray: 90 positions for primary or
secondary tubes
- Sample volume: 2~45ul, step by 0.1ul
- Sample probe: liquid level detection, clot detection and collision
protection
- Probe cleaning: automatic washing for both inside and outside; carry over
< 0.1%
- Automatic sample dilution: pre-dilution and post-dilution; dilution ratio
up to 150; dilution vessel: cuvette
Internal bar code reader: - Used
for sample and reagent programming, Capable to link in LIS bi-directional mode
- Applicable to various bar code systems of code 128, code 39, code 93, I
2of 5, codabar, ITF, UPC/EAN
ISE Module: - Optional selection for
K+, Na+, Cl- or K+, Na+,
Cl-, Li+ - Throughput: Up to 240 or 320 tests per hour
- Reagent Handling: - Reagent tray: 80 positions in refrigerated
compartment (4~10°C)
- Reagent volume: R1: 150~350ul, step by 1ul, R2/R3/R4: 20~350ul, step by
1ul 2 independent reagent probes - Reagent probe: liquid level detection
and collision protection, inventory checking
- Probe cleaning: automatic washing for both inside or outside, carry over
< 0.1%
Reaction System: - Reaction rotor: rotating tray, 90
cuvettes with automatic washing, Independent mixing probe
- Cuvette: optical length 5mm, Reaction volume: 150~360ul
- Operating temperature: 37°C ± 0.1°C
Optical
System: - Light Source: Halogen-tungsten lamp
- Photometer: reversed optics, grating photometry - Wavelength: 340,
380, 412, 450, 505, 546, 570, 605, 660, 700, 740, 800nm Resolution: 0.001Abs
- Absorption range: 0~3Abs (10mm conversion) - Control and Calibration:
- Calibration mode: linear (one point, two points and multi-points),
Logit-Log 4P, Logit-Log 5P, spline, exponential, polynomial, parabola -
Control software: Westgard multi-rule, Cumulative sum check, twin plot
Operation Unit:
- Operation system: Windows 2000 Professional or Windows XP -
Interface: RS-232 - Working Conditions: - Power Supply: 200-240V~,
50/60Hz, 1500 VA
- Or 110-130V~, 50/60Hz, 1500 VA
- Temperature: 15-30℃ - Humidity: 35-80%
- Water consumption: <20L/hour De-ionized water - Dimension:
1180mm×700mm×1145mm (W×D×H) - Weight: 300kg
|