HomeHome
  Company ProfileDiagnosticsDownloadsContact
DocumentationInstrumentsOrder InformationCustomer Login
NewsOEMTraining
 
 
 
Autolyser
Autolyser 400
Photometer DTN405
Photometer DTN-410K
Photometer DTN-410
DIAcheck WB for whole blood
DIAcheck C1
DIAcheck C2
DIAcheck C4
DIAcheck CP4
CoaLyser
ELISA Reader
ELISA Washer
Strip Reader 40
Strip Analyser 500
Photometer DP 300
Photometer DP 310
DIAminiLAB
 
 
Instruments > ELISA Reader
 
ELISA MICROPLATE READER (DIAREADER)


MODELS:

  • DIAREADER ELX800G:
measuring range from 400 nm to 750 nm
for reading 24, 48 or 96-well plates

  • DIAREADER ELX800 UV:
extended measuring range from 340 nm to 750 nm
for reading 24, 48 or 96-well plates

  • DIAREADER ELX800NB:
measuring range from 400 nm to 750 nm
for reading 384-well microplates as well as 24, 48 and 96-well plates


TECHNICAL SPECIFICATIONS


  • Microplates: 24, 48, 96 - well microplates(model ELX800NB also 384 wells)
  • Speed: 30 seconds for 96 wells
  • Wavelength range: 400-750nm (model ELX800UV 340 nm - 750 nm )
  • Filters: 5-cavity wheel,
  • Absorbance range: 0.000-3000 Abs
  • Resolution: 0.001 Abs
  • Accuracy: +/-1% +/- 0.0010 Abs from 0 to 2.000 Abs
  • Linearity: +/-1% from 0 to 2.000 Abs, +/-3% from 2.000 to 3.000 Abs
  • Repeatability: +/- 0,5% +/- 0.005 Abs from 0 to 2.000 Abs
  • Software Interfacing: Gen5 (TM)

  • On-board data analysis:
Simply to use and programm via the reader "soft-keys" and 2x24 LCD display, the DIAREADER's on-board data reduction capabilities surpasses that of most PC software packages:
- Extensive curve fitting: linear, cubic, quadratic, 4-P, log-logit, cubic-spline, point to point
- Auto or manual plate layout for blanks, controls, standards and samples
- Control and assay validation
- Transformation and Formula Applications
- Cut-off and Call Criteria
- Up to 75 assay programmed into DIAREADER's memory which can be recalled instantly
- Up to 8 microplates test results saved in the readers's memory
- Panel assay capable
- Store 25 standard curves

OPTIONS:
  • Universal test plate
- N IST traceable
- Checks alignment, accuracy, reproducibility and linearity
- Confirm PowerWave xTM and µQuantTM wavelength accuracy with 7-filter test plate
-Use with KCjuniorTM for complete GLP compliance

  • Gen5(TM): Basic data analysis and curve fitting
- 32bit WindowsTM application controls all DIALAB Absorbance Readers
- Intuitive and flexible protocol creation
- Consise data analysis and reporting
- Flexible curve fitting options
- Contex-sensitive help screens

  • Printer: HPTM or EPSONTM compatible printer

ELECTRICAL DATA:

  • Power: External, 24 volt power supply running from 100-240 VAC, +/- 10%, 50-60 Hz
  • External outputs: RS 232 bidirectional serial interface, Parallel Centronics printer interface
  • Display: 2x24 LCD display
  • Regulatory: CE Mark, ETL listed, conforms to UL 3101-1, conforms to IEC 1010-1, certified to CAN/CSA C22.2 No. 1010.1


PHYSICAL DATA

  • Dimensions: 41,9x38,1x17,8 cm (DxWxH)
  • Weight: 8,4 kg