Smallest footprint on the market for a fully automated 8 pin wash head ELISA system
Learned error recovery: Supporting walk-away automation with trained error recovery actions
LIS and LIMS compatible
QC monitoring
Sample identification with an optional on-board barcode reader
Compatible with DIALAB® ELISA assays
No. of plates: 2
No. of assays: 1 assay per strip or up to 12 assays/plate
No. of samples: 100
No. of reagents: 18
No. of Std/Ctrl: 24
No. of pipettes: 1
Reader
Washer
Incubator
Sample module
Ambient drawer
Barcode reader (optional)
Liquid tanks for system- and waste water
Sample tip size: 300 μLSample pipetting volume: 10 – 250 μL
Precision: ≤ 3% CV (10 - 250 μL)
Dilution range: 1 to 5000
Replicates: Up to 96 samples, standards and controls
Number of sample-tips: 2 racks of 108
Sample tube dimensions: 10 – 16 mm Ø, 40 -100 mm length
Reagent tip size: 1300 μL
Number of reagent tips: 20
Reagent pipetting volume: 20 – 1000 μL
Reagent pipetting precision: ≤ 3% CV (10 - 1000 μL)
Manifold configuration: 8 – way
Programmable volumes: 50 – 1000 μL
Wash containers: 2 wash bottles at 2.0 L, with level-sensing
Waste containers: 1.5 L with waste-full sensor
Residual wash volume: < 3 μL
Number of incubators: 1
Temperature range: RT, 4° C to 40° C
Temperature accuracy: ± 1° C
Shaking: 14-20 Hz periodic or continuous
Photometric range: 0.000 – 3.000 OD
Spectral range: 405 nm – 690 nm
Filter: 405 nm, 450 nm, 490 nm, 620 nm
Precision: < 1% CV (< 2.000 OD); < 2% CV (2.000 – 3.000 OD)
Accuracy: 0.005 OD or 2.5%
Reading time: < 30 seconds, single wavelength
Power supply: 100 – 240 VAC, 50/60 Hz
Power consumption: < 300 VA, UPS recommended
Net weight: 45 kg
Gross weight: 80 kg
Dimension: 660 x 540 x 680 mm (H x W x D)