Throughput up to 420 tests/h without I.S.E (up to 600 tests/h with I.S.E)
Optional with I.S.E module (Na, K, Cl, CO2)
Test modes: Routine, Batch, Emergencies (STAT), Profiles
Up to 500 tests in memory+ unlimited relation tests
Sample: Serum, Urine, Plasma, CSF
Sample & Calibrator pre-dilution
Cooled reagent tray
Cooled calibrator & control tray
Non disposable optical glass cuvettes
Solid state photometer with 14 channels 340-800nm
Clot sensor
Crash sensor
QC program
Internal barcode reader for sample and reagents
Bidirectional LIS
No existing water and waste lines necessary
Reagent tray: 80 positions, cooled
Sample tray: 100 positions
Calibrator & control tray: 26 positions, cooled
Reaction tray: 84 glas cuvettes, heated
Sample volume: 1.5 μL to 100 μL
Reaction volume: 180 μL to 400 μL
Cuvette temperature: OFF/37°C ± 0.2°C (Peltier controlled)
Solid state photometer: 14 channel + 1 ref. channel
Wavelengths: 340, 380, 405, 436, 478, 510, 546, 578, 600, 630, 660, 700, 740, 800 nm
Absorbance range: 0.000 – 3.000 Abs
Sensitivity: ± 0.001 Abs
Accuracy: ± 1% from 0 to 2.000 O.D
Tungsten halogen lamp with dichroic reflector
Lamp Life time approximately 2000 hours
1x Sample pipetting arm
2x Reagents pipetting arms
3x Long life ceramic dilutors
Dilutor precision: ± 0.1 %
Dilutor resolution ± 0.1 μL for serum
± 1.0 μL for reagents
Power supply: 100 / 240 VAC, 50-60 Hz, 550 Watt
U.P.S.: 1500 Watt
Net weight: 162 kg
Gross weight: 237 kg
Dimensions: 1150 x 1150 x 720 mm (H x W x D)
With an access you are able to visit the areas which are available for registered customers only.